Benchtop Autoclave Steam Sterilizer 20L 24L Fast Speed Sterilizing
Model: MCS-XB20J
Charcteristucs:
1. Indicator light indicates working state.
2. For 4~6 minutes rapidly sterilizing.
3. Sterilizing temperature and time can be preset.
4. Steam-water inner circulation system: no steam discharge,and the environment for sterilizing will be clean and dry.
5. Over temperature & over pressure auto-protection.
6. Safe protection of water lacking.
7. Exhaust the cool air automatically.
8. With three stainless steel sterilizing plates.
9. The chamber of the sterilizer is made of stainless steel.
10. Automatically shut off with beep reminding after sterilization.
Technical data:
MODEL | MCS-XB20J | MCS-XB24J |
Sterilizing chamber volume | 20L(φ250×420 mm) | 24L(φ250×520 mm) |
Maximum working pressure | 0.22Mpa | |
Maximum working temperature | 134℃ | |
Adjustment of temperature | 105-134℃ | |
Timer | 0-60 min | |
Chamber temperature equal | ≤ ± 1℃ | |
Source power | 1.5KW/AC 220V 50Hz (AC110V 60HZ) | |
Sterilizing plate | 340×200×30 mm (3 piece) | 400×200×30 mm (3 piece) |
Dimension | 480×480×384 mm | 580×480×384 mm |
Package dimension | 700×580×500 mm | 800×580×500 mm |
G. W/N.W | 44/40 kg | 50/45 kg |
The working environment and condition of the product:
1) environment temperature: 5~40℃
2) Relative humidity: ≤80%
3) Atmospheric pressure range: 86~106 KPa
4) Power supply:
AC220V±11V, 50Hz±1Hz ( should have ground lead)
Transport and storage environment and condition:
1) environment temperature: -10~50℃
2) Relative humidity: 10%~90%
3) Atmospheric pressure range: 50~106 KPa
With over-temperature, over-pressure auto-protect device:
1) This product equipped with safety valve, and will auto release when steam pressure over 34Psi(0.24MPa).
2) The power will be cut -off automatically, the water be cut off and alarm when the heating tube temperature over 150 ℃ in the course of drying.
Working principle
The sterilizer is designed by using gravity type pressurized steam sterilization principle. Put 0.6 liter of distilled water into sterilization chamber, use electric heater to heat directly to produce pressurized steam. The pressurized steam act with microorganism on the objects to be sterilized to destroy its structure and reach the purpose of sterilization.
Installment & Adjustment
l) The Sterilizer should be put on a horizontal working table with all four supporting feet contacting the table surface evenly,
2) Adjust two front adjustable supporting feet to make the angle of axis line of the sterilizer and the table surface about 4º, so that the 0.6 liter water can be fill into the tank and objects to be sterilized could be put in the instrument plate stably and the discharge of the leftover water in sterilization chamber could be thorough and clean (See Figure 3);
3) Check to see whether the power supply condition conforms to the product specifications.
OPERATION of the steam sterilization:
l ) The preparation and inspection before use
1.l Put the sterilizer stably on a horizontal working table, adjust the adjustable supporting feet, make the angle of axis line of the sterilizer and the table surface about 4º
l.2 Check power supply
l .2. l Power supply data conform to product requirement. Single phase (AC220V, 50Hz).
l .2,2 Connecting to the power supply
l.3 Make a reliable earth connection.
1.4 Prepare 4 liters of distilled water.
l.5 The binding of dressing and textile can not be too tight
1.6 Certain amount of sterilization indicator (chemical indicating piece or biologic indicating piece)'
1.7 There should be water containers under the overflow and drain outlet of the sterilizer.